HTM 2022/02-01

MONITORING AND ALARM SYSTEM

GENERAL DESCRIPTION


Delta P can supply alarm systems manufactured in accordance with HTM2022/02-01 memorandum.

In particular the following models are available:

 

  • Area alarm panel HTM 2022/02-01
  • Plant alarm system HTM 2022/02-01

 

Area alarm panel HTM 2022/02-01 General description

The area alarm panel is used to monitor pipeline pressure within a ward, theatre etc.

The pressure is monitored by pressure switches in the pipeline downstream from the last AVSU. Pressure is monitored for both high and low pressure (low only for vacuum

 

Plant alarm system HTM 2022/02-01 General description

The Plant Alarm System is intended to be used as central indicator panel in order to display warning and alarm conditions for medical gas supply systems. It must be located in a position where there is continuous 24-hour occupation, such as the telephone switchboard room or the porter’s lodge.

MAIN FEATURES


AREA ALARM PANEL HTM 2022/02-01

  • It allows the monitoring of both high and low pressure (low only for vacuum) up to 6 gases contemporary.
  • Alarm system fault in accordance to HTM 2022 / 02-01 requirements
  • Possibility to check of integrity of the cabling between alarm system and pressure switches with a termination box supplied on request

 

  • Designed to meet the requirements of:
  • HTM 02-01 / 2022 Medical Gas Pipeline Systems
  • C11 NHS Model Engineering Specification – Medical Gases
  • BS EN ISO 7396-1 Pipeline systems for compressed medical gases and vacuum
  • 2004/108/EC The EMC Directive
  • 2006/95/EC The Low Voltage Directive

 

PLANT ALARM SYSTEM HTM 2022/02-01

  • It allows the monitoring of the status of the supply unit from 1 to 6 gases contemporary, giving up to 4 alarm conditions plus normal for each gas.
  • Any service may be split into four individual displays to be used to bring signals, such as the common alarm output from Area alarms, to a central point or to any point where this information is required
  • Alarm system fault in accordance to HTM 2022 / 02-01 requirements
  • The battery reserve enables the system to remain in operation for up to 8 hours
  • Remote transmitters available on request

 

  • Designed to meet the requirements of:
  • HTM 02-01 / 2022 Medical Gas Pipeline Systems
  • C11 NHS Model Engineering Specification – Medical Gases
  • BS EN ISO 7396-1 Pipeline systems for compressed medical gases and vacuum
  • 2004/108/EC The EMC Directive
  • 2006/95/EC The Low Voltage Directive